Flow cytometry
From DrugPedia: A Wikipedia for Drug discovery
Current revision
== Flow Cytometry == Introduction Flow cytometry uses the principles of light scattering , light excitation, and emission of fluorochrome molecules to generate specific multi-parameter data from particles and cells in the size range of 0.5um to 40um diameter. Cells are hydro-dynamically focused in a sheath of PBS before intercepting an optimally focused light source. Lasers are most often used as a light source in flow cytometry. Flow cytometers use the principle of hydrodynamic focusing for presenting cells to a laser (or any other light excitation source). The sample is injected into the center of a sheath flow. The combined flow is reduced in diameter , forcing the cell into the center of the stream. This the laser one cell at a time. This schematic of the flow chamber in relation to the laser beam in the sensing area.(From Current Protocols In Cytometry, Unit 1.2 , p1.2.2 ).
As your cells or particles of interest intercept the light source they scatter light and fluorochromes are excited to a higher energy state. This energy is released as a photon of light with specific spectral properties unique to different fluorochromes.
One unique feature of flow cytometry is that it measures fluorescence per cell or particle. This contrasts with spectrophotometry in which the percent absorption and transmission of specific wavelenths of light is measured for a bulk volume of sample.Fluorescence spectra of commonly used fluorochromes. Scattered and emitted light from cells and particles are converted to electrical pulses by optical detectors. Collimated (parallel light waveforms) light is picked up by confocal lenses focused at the intersection point of cells and the light source. Light is send to different detectors by using optical filters. For example , a 525 nm band pass filter placed in the light path prior to the detector will only allow “green” light into the detector. The most common type of detector used in flow cytometry is the photomultiplier tube (PMT).The electrical pulses originating from light detected by the PMTs are then processed by a series of linear and log amplifiers. Logarithmic amplification is most often used to measure fluorescence in cells. This type of amplification expands the scale for weak signals and compresses the scale for “strong” or specific fluorescence signals.
After the different signals or pulses are amplified they are processed by an Analog to Digital Converter (ADC) which in turn allows for events to be plotted on a graphical scale(One Parameter, Two parameter Histograms).
Flow cytometry data outputs are stored in the form of computer files using the FCS 2.0 or 3.0 standard. Data corresponding to one sample can be stored as a listmode file and/or histogram file.
Histogram Files Histogram files can be in the form of one-parameter or two-parameter files.Histogram files consist ofa list of the events corresponding to the graphical display specified in your acquisition Protocol.One-parameter histograms A one-parameter histogram is a graph of cell count on the y-axis and the measurement parameter on x-axis. All one-parameter histograms have 1,024 channels. These channels correspond to the original voltage generated by a specific "light" event detected by the PMT detector. In other words, the ADC assigns a channel number based on the pulse height for individual events. Therefore, brighter specific fluorescence events will yield a higher pulse height and thus a higher channel number when displayed as a histogram.
Two-Parameter Histograms A graph representing two measurement parameters, on the x- and y-axes, and cell count height on a density gradient. This is similar to a topographical map. You can select 64 or 256 channels on each axis of two-parameter histograms. Particle counts are shown by dot density or by contour plot.Two parameter histogram Dot Plot displaying FL1-FITC on the x axis and FL2-PE on the y axis.
Listmode Data Files Listmode files consist of a complete listing of all events corresponding to all the parameters collected, as specified by your acquisition Protocol. This file follows a format specified by the FCS 3.0 standard. Raw listmode data files can be opened or replayed using any program designed for analysis of flow cytometry data. You should keep in mind that a Protocol serves as a template. It allows you to collect specified Parameters (i.e. FLS, FL1, FL2, etc.), and how these parameters are displayed. Protocols also serve to determine how the data is Gated, and contains all the Regions from which your statistics will be generated. In addition, Protocols contain other specific information that serves as direct interface between the computer workstation and the cytometer. These pertain to high voltage settings for the PMT detectors, gains for amplification of linear parameters, sample flow rates, fluorescence compensation, discrimination settings, etc.
Once your data has been collected and written into a listmode file you can replay the file either using the specific Protocol used for collection or any other program specifically designed for analysis of flow cytometry data. However, you should keep in mind that you can only adjust Regions, Gating, and Parameters to be displayed. Settings such as amplification, fluorescence compensation, etc., can not be modified . Therefore, when collecting data make sure that your instrument settings are correct. Finally, if you open your listmode files using a programs such as FlowJo, WINMIDI, and/or ExPO you will have to specify parameter displays, and create Regions and Gating corresponding to the Protocol used for collecting the data.Flow Cytometry Analysis and Sorting Flow cytometry analysis of a single cell suspension yields multiparameter data corresponding to Forward Light Scatter (FLS), 90° Light Scatter (90LS), and FL1-FL4. This information allows researchers to identify and characterize various subpopulations of cells. The process of separating cells using flow cytometry multiparameter data, is referred to as sorting.
The Beckman-Coulter XL instruments are bench-top, flow cytometer, analyzers. They are capable of acquiring multiparameter flow cytometry data but they can not separate or purify cells. Sorting is a specialized process that requires sophisticated electronic components not incorporated into most bench-top instruments.
The Beckman-Coulter ELITE-ESP and Dako-Cyomation MoFli high speed sorters are representative of a research grade cell sorter. Sorters include the following components:
1) A tunable transducer which permits the breaking of the fluid sheath into individual droplets. These individual droplets will encapsulate single cells.
2) Electric charge delays for charging individual droplets.
3) Deflection plates for deflecting individually charged droplets into collection tubes.
4) Software settings for defining sorting criteria, these include regions defining populations to be sorted.